Showing all 1 results
Compliant with DIN13620-2 and IS09170-1 standard
Gas Type: O2, MEDICAL AIR, VAC, N2O, N2, CO2 are available
Available in wall type with box, console type on bedhead units and hose barb type
Secondary isolating valve enables easy maintenance, without shutting down the pipeline system
Color-coded for easy gas identification according to ISO32
100% pressure and leakage tested
Second-pull-out design can prevent the user from accidentally pull out the probe
15000 times Swap Fatigue Test
Suitable ambient temperature: -20℃ to 60℃, ambient humidity: 10% to 95%
The working pressure for oxygen, air (breathing), nitrous oxide, and carbon dioxide is 0.4MPa
The working pressure for nitrogen and instrument air is 0.8MPa
The negative pressure working pressure is -0.075MPa
The maximum gas supply flow rate is more than 100L/min; the maximum suction flow rate is more than 40L/min